Skip to main content

New resources available to accelerate the adoption of novel digital clinical measures in clinical research

Take the fast track to becoming a V3 and V3+ expert

V3 in Action_Koneksa

“The first step [of V3] is sensor verification, and to understand the accuracy of accelerometers of [three devices], we put them on top of the Shake Table, which is the gold standard benchtop device for evaluating sensor accuracy…[we determined that] all three devices are fit-for-purpose to record gait and balance… [When looking at] hand tremors, another feature of Parkinson’s Disease, the accuracies are different… if we didn’t do this verification and we wanted to use these devices for collecting hand tremors it shows that this solution is not optimal.” – Elena Izmailova, Chief Scientific Officer, Koneksa

Learn more about this V3 in Action story: